Sample and Standard peak area found in the ratio of 1:6 ratio during HPLC analysis

Image
 Sample peak at RT about 4.5 min. with height of about 90 Standard peak at RT about 4.5 min. with height of about 630 Investigation : When such phenomenon was observed then analyst was interviewed and detailed discussion done about level of dilutions done during sample and standard solution preparation. As per the Standard testing procedure, std was to be taken as 50 mg and sample was to taken as Equivalent to 50 mg active substance. Now, the analyst weighed correctly 50.2 mg of standard but during sample weighing he took the similar weight as 50.5 mg. Here the point is weight of tablet under analysis is 300 mg. Hence, the actual weight is to be taken as 300 mg as each tablet contains 50 mg of active content. This resulted in Sample peak height to Standard peak height in the ratio of 1:6. Conclusion : This is the common mistake in pharma lab which is sometime done by analyst. They should clearly distinguish between the word Equivalent and Equal . Though this seems silly mistake b...

Drift in the analyte Retention time in HPLC analysis

Image
Drift in Retention time of Analyte peak during HPLC analysis: Causes:  There are several reasons of observing drift in Analyte peak in HPLC analysis  like: a. There is slight leakage in pump due to which RT shift is observed in the chromatogram b. Insufficient  saturation of HPLC column as every column behaves in different manner and take different time to get saturated c. Insufficient sonication of Mobile phase which involves solvents like Acetonitrile which requires minimum 15 to 20 mins. of sonication to remove entrapped air bubbles d. Incorrect selection of column for analysis. Remedies :  a. Any type of leakage which might at pump or detector, must be immediately fixed. It might be possible fer rules are not tightened properly which must properly tightened  b. Column saturation  should be done to get stable baseline. Once it is free from any types of noise, standard injection to be done. c. Mobile sonication should be done for minimum 20 minutes to avo...

Peak of Lanreotide not detected on Thermo HPLC

Image
Description:   Peak of Lanreotide not detected on HPLC of Thermo make. Investigation:  Investigation was done to find the root cause for above issue. Mobile phase changed multiple times. All HPLC tubings were checked for any kind of leakage but no issue was observed. Root Cause : To identify the root cause service engineer of Thermo was contacted. He took multiple trails but not succeeded.  After that he identifies that during method preparation calibration curve value taken was 5. As per his opinion it should be 6 for Thermo HPLC otherwise principle will not be detected. The same method was previously run on waters make HPLC where calibration curve was 6 and the same wss latter included in method, but when run on Thermo make HPLC,  peak was not detected. Conclusion: It is to be noted that calibration curve value play significant role during run on different HPLC systems. It is to be modified during development studies so that final method should include the correct ...

Interference of Toluene peak during GC Analysis

Image
  Description of Incident:  During analysis of one of the Active content during GC analysis, peak interference of Toluene observed.  Investigation:  During Investigation it was found that in every injection on GC, peak of Toluene was eluting. It was checked and found that Toluene was not used in that particular analysis then also Toluene peak was eluting. To rule out the possibility of chemical contamination, new bottle of DMSO was opened. After using new bottle reagent also, Toluene peak was eluting  and no significant drop in area was observed. Root cause:  During further Investigation  it was observed that during previous analysis Toluene was used as an standard.  Currently the oven temperature  was kept as 120°C (as boiling point of Toluene is 110°C) to remove the traces of Toluene but then also Toluene peak was observed.  Remedy :  As a corrective action, oven temperature was increased to about 200°C and column condition done f...

Peak area observed less than the trend during HPLC analysis

Image
Trouble description : During HPLC analysis, sometimes it happens that there is significantly less area observed as compared to previous trend od peak area. Investigation :  During investigation everything was checked i.e.: Checke Solution dilutions prepared for sample and standard and found Ok. Checked Column leakage and found no leakage. On changing the HPLC system, same solutions were used and arwawere checked  Mobile phase filteration checked and it was found that analyst did not filterd the mobile phase As unfiltered mobile phase was used,so solvent inlet filter was checked On inevstigating it was found that they were completely choked. Action taken : Solvent inlet filter was changed and analysis performed. Area observed in trend. CAPA: Solvent filter should be sonicated for 30 mins if any variation is observed during the analyis.

XZ drive failure in Ion Chromatography of Thermo Scientific make

Image
  In an reputed pharma company, in IC system, XZ drive failure was observed.  Root cause analysis:  The system was shutdown for about 8 months. So initially it was not known that why such thing happened. Just as instrument was started and XZ drive failure took place. During investigation it was found that earthing observed as 2.4 ohms which is more than the requirement of NMT 1 ohms. Actually at the time of installation it was 1.4 ohms two years back. Due to increase in load, earthing increased upto 2.4 ohms which resulted in failure of XZ drive of IC instrument.  Cost of new drive is about $ 9000. Precaution: It is highly advisable to keep close check on earthing during operation of sophisticated instruments like IC system otherwise it can result in high costly affair.

Case study of Incident observed during HPLC analysis

Image
Case study of Incident observed during HPLC analysis on Waters auto sampler:   Recently during HPLC analysis on Waters make HPLC, a very unique incident happened. During Related substance analysis, after 3 standard injection, in fourth injection missing vial was observed though injection was taken from same single vial. Root cause analysis: Further on investigation it was identified that at the time of vial labelling, analyst used small sticker. So at the time of vial picking by robot in autosampler, small piece of sticker on inside wall of vial carousol. Due to this robot tried to pick the vial but due to resistance from vial, robot was not able to pick the vial. This resulted in above incident and missing vial message was obseeved. On prima facie it was difficult to find the root cause but due to so small sticker, such sticker can cause the incident. Remedy :  Never apply sticker labels on vials and always use marker pen to assign the identification on vial. Hope you may get...

Oh!! Ghost peaks during HPLC analysis

Image
Though this is funny word 'Ghost' but after all this is also one of the major issue during HPLC analysis. There are many factors which gives rise to such unidentified peaks. Some of important factors are mentioned below: 1) Late eluting peaks: During analysis many times late eluting peaks are found in which they interfere in principle peak in subsequent injections. This can be solved by increasing the run time. 2) Contamination: It sometimes happens that due to some contamination ghost peak used to found during analysis. To handle such situations Ghost buster columns are available in market which can be used. It looks like shown in below image: 3) Water Quality: While preparing mobile phase water quality is not proper then it may give rise to ghost peaks. Such situation should be handled by using either Milli Q grade water or HPLC grade water. 

Peak splitting during HPLC analysis

Image
Many times peak splitting is observed during HPLC analysis. Majorly there are 3 main impacting factors due to which peak splitting is observed. 1) Contaminated guard column: Usually guard columns are not washed and this gives splitting of peaks when used for long time. This can be solved by reversing the guard column and flushing it with solvents. 2) Partially blocked frits:  This does not come to notice easily but is very important that frits should be timely changed or cleaned. Since coagulated frits will give rise to splitting of peaks. 3) Small uneven void at column inlet: Repack column with the particles or column should be flushed by reversing it. This will definitely give rid of peak splitting.

Hey!!! No peak found during HPLC Analysis

Image
Many times it happens that no peak is found during HPLC analysis. There are many such factors which gives rise to such situation. Out of that we will talk about 4 major factors today: 1) Solvent issue:   Practically during analysis many times it happens that analyst uses wrong solvents for example instead of us in ACN they use Methanol. Due to this change in mobile phase, peak is not eluted. So very first it should be checked that correct solvent is used during HPLC analysis. 2) Detector Response:   Second important factor is detector response. If there is any issue in detector response peak will not elute. Ensure that at the time if peak is not eluting then check the detector for any type of error. 3) Injector issue:  Third major issue is with Injector. There can be any bubble entrapment which may give low response or no peak during HPLC analysis. Hence system should be purged properly to avoid such incidences. 4) Leak issue:   Leak is also one of the most impacting...

3 impacting factors for peak broadning during HPLC analyais

Image
During HPLC analysis many times a issue is observed namely peak broadening. Due to this area increases abnormally giving rise to erroneous results.  Some of the reasons and solutions for peak broadening are given below : 1) Temperature: Ambient temperature or room temperature can impact the broadening of peaks. To control it, column temperature should be kept in equilibrium mode using column heater or column cooler. 2) Equilibrium of column: Column and detector should be stabilized properly. During gradient system, sufficient time should be allowed between analyses and re-equilibriation.  3) Incorrect column: Check that whether correct column is used during analyses with respect to dimensions as well as packing. If incorrect column is used it will give broad and improper peaks.

5 causes and solution for changing retention time.

Image
There are many factors impacting Retention time during HPLC analysis. Among them 5 most impacting parameters and there solutions are as below: 1) Contamination build up: This is related to some contaminants which gets retained in the column. This can be solved by continuous flushing using strong solvents. This will flush out the residue and constant retention time will be achieved. 2) Equilibration time: If equilibration time for gradient is not sufficient then it will give variation in retention time. The equilibration time should be sufficient so that constant retention time is achieved. 3) Online degassing: If online degassing is not proper then it will affect the retention time of the peak. Online mobile phase degassing is done by the in build degassed on HPLC system which should be checked from time to time. 4) Evaporation : If reservoir bottles of mobile phase and solvents are not covered properly, slow evaporation of solvent will cause change in mobile phase composition. This ...

4 causes and solutions for %RSD Failure in HPLC analysis

There are multiple reasons of %RSD failure during HPLC analysis but four are the major factors impacting %RSD: 1) Degassing of mobile phase: There are some solvents which requires proper degassing. So after preparation of mobile phase, it should be properly sonicated. If it is not properly degassed it will definelty result in variation of peak area which will finally give RSD failures. 2) Standard stability: There are certain molecules which are very sensitive to temperature. So if run time is long then till they reach up to 5th or 6th replicate, there area drops drastically which results in RSD failure. Hence method should be developed in such a manner that such molecules should be taken care of. 3) Pressure drop: Due to pressure drop in chromatograms it will impact the area of standard. Hence again this can be attributed to improper degassing of mobile phase which results in %RSD failures. 4) Improper priming of HPLC system : Before analysis, all solvent lines should be prime pro...

Let's talk about ODS and BDS Column in HPLC

Image
Definitions: ODS column: Full form of ODS column is Octadecasilane because it contains Octadecasilane chain in it. This column is mainly used in reverse phase chromatography and has free -OH group. Also, as a trade name it is also called as Inertsil column. Using this column usually gives less tailing in peak.  Pros : The main advantage of these columns is it's low cost and sharpness of peaks. Con's: The main disadvantage of using these column is they take too long for column saturation. Elution of peak is fast and so it becomes difficult to separate certain components. BDS Column: The full form of BDS is Base deactivated silica. This column is mainly used in reverse phase chromatography but has blocked -OH group. These column is also called End capped columns. Also, its trade name is Hypersil column. Using this column gives too high tailing in peak. Pros: The main advantage of this column is that it is used for analysis of basic compounds.   Con's: The main disadvant...

3 ways to get rid of %Carryover in HPLC system

 %Carryover in HPLC system is the residue which interferes during routine analysis on HPLC which impacts the actual results. This is one of the most common problem which every analyst encounters during HPLC analysis. Though as a part of system limitations there are certain acceptance criteria for % Carryover i.e. it should not be more than 0.1% of the principle peak, as a common practice. It is very easy to get rid of such %carryover by following three (3) methods: 1) Injection pattern: During analysis injection pattern matters a lot. If high concentration solution is injected first followed by low concentration silution then %carryover is more. So if any high concentration solution is injected it should be followed by blank injection to avoid any %carryover. This is acceptable even by the auditors if more blanks are added. This can also avoid creation of any type of lab events or incidences. 2) Frit cleaning: There is a part called frit after which solution pass into the detector...

5 Major causes of noise in chromatograms during HPLC analysis?

There are 5 major causes for too much noise during HPLC analysis? Noise is the basic problem which every analyst faces during day to day HPLC analysis. Below are the causes which create noise and if properly handled, will definitely solve this problem. 1) Sampling rate: If too much noise is observed then very first check the sampling rate in HPLC method. It should be 1.0 instead of 10 (preferably in case of waters HPLC systems). 2) Pressure graph: Pressure graph to be monitored. If pressure fluctuations are more then it will give rose to noise creations during analysis. 3) Degassing: Degassing of mobile phase. If mobile phase is not degassed properly then it will take time to stabilize the column which will again generate the noise. 4) Detector: If room temperature is fluctuating too much then it will impact on detector which will disturb the baseline and will result in disturbed baseline.  5) Column saturation: Some mobile phase requires some long time to saturate during HPLC a...